U S E R M A N U A LMODEL SERIES: PM3000 PM3100 (shown) PM3200 PM3500 PM3600SAVE THESE INSTRUCTIONSCAUTIONFederal (USA) law restricts this device
9DISPOSAL INSTRUCTIONSDispose of the Vacuum Regulator in accordance with the local regulations.Please RecycleWARNINGProduct should be cleaned before b
10LIMITED WARRANTY AND LIMITATION OF LIABILITYPrecision Medical, Inc. warrants that the Medical Vacuum Regulator (the Product) will be free of defects
DECLARATION OF CONFORMITY Precision Medical, Inc 300 Held Drive Northampton PA 18067, USA Emergo Europe (European Office) Molenstraat 15 2513 BH
1CONTENTS RECEIVING / INSPECTION ... 2INTENDED USE ...
2RECEIVING / INSPECTIONRemove the Precision Medical, Inc. Vacuum Regulator from the packaging and inspect for damage. If there is any damage, DO NOT U
3SPECIFICATIONS GAUGE RANGE: *Counterclockwise direction PM3000: 0 - 200 mmHg - Full Vacuum PM3100HV: 0 - 300 mmHg - Full Vacuum*PM3000E: 0 - 200 mm
4OPERATING INSTRUCTIONSCAUTIONInspect the Vacuum Regulator for visual damage before use,DO NOT USE if damaged.NOTE: Overflow protection should be used
5PARTS DESCRIPTIONCAUTIONMissing or illegible labels must be replaced, contact Precision Medical, Inc.49875612Selector KnobRegulator Knob1 QUANTITY -
6REPAIR KITSAnalog Part# Digital Part#PM3000 / PM3000D Vac Reg RK6000 RK6000DPM3000E / PM3000DE Vac Reg RK6000E RK6000DEPM3000HV / PM3000DHV Vac Reg R
7VACUUM REGULATOR CLEANING ILLUSTRATION(Cleaning/Decontamination Instructions on page 8)Suction TubingColdDisinfection/Sterilization SolutionPatient P
8CLEANING / DECONTAMINATION(As needed)1. Attach a working Vacuum Regulator with a continuous regulated mode to a minimum vacuum source of 15 inHg.2. M